Multichrom

Multicapillary column applications.

   Multichrom, Ltd

Applications of multicapillary columns in laboratory gas chromatographs 


Download file (pdf - 232Kb)


Analysis of solvents

Analysis of solvents
 
Chromatograph

Detector
Injection
Column
Length
 Capillaries
LSP
 Temperature
Carrier gas
  TSVET-500M,
upgraded
FID
Splitting 1:1
Coiled MCC
1 m
40 μm, 103 
OV-1701, 0.2 μm
500C
Ar, 80 ml/min
Peaks:
1 – acetone;
2 – ethyl acetate;
3 – benzene;
4 – n-butanol;
5 – toluene;
6 – ethylcellosolv;
7 – butyl acetate;
8 – ethylbenzene;
9 – m-xylene;
10- o-xylene;
11- styrene;
12- cyclohexanone;
13 – cyclohexanol.



Control of drugs and narcotic substances

Control of drugs and narcotic substances
Opium extract
 

Chromatograph
Detector 
Injection
Column
Length
 Capillaries
LSP
 Temperature
Carrier gas

  Kristall 2000M
FID
Splitting 1:1
Coiled MCC
1 m
40 μm, 103
OV-1, 0.2 μm
2000C
Ar, 89 ml/min
Peaks:
1 – methylstearate (standard);
2 – codeine;
3 – morphine;
4 – tebaine;
5 – papaverine;
6 – narcotine



Control of drinking water quality.
Analysis of head space vapors


Control of drinking water quality

Volatile halogenated hydrocarbons in water
 

Chromatograph

Detector
Injection
Column
Length
 Capillaries
LSP
 Temperature
Carrier gas

 

Tsvet-500M,
upgraded
ECD
0.5 ml; splitting 1:1
Coiled MCC
1 m
40 μm, 103
OV-1301, 0.4 μm
400C
Ar, 65 ml/min

Peaks:
1 - 1,2-dichloroethane;
2 – chloroform;
3 - carbon tetrachloride;
4 - trichloroethylene;
5 – dichlorobromomethane;
6 - tetrachloroethylene;
7 - dibromochloromethane;
8 – bromoform.



Control of pesticide contaminations

Control of pesticide contaminations

Chloroorganic pesticides.
 

Chromatograph

Detector
Injection
Column
Length
 Capillaries
LSP
 Temperature

 Carrier gas

  TSVET-500M,
upgraded
ECD
Splitting 1:1
Coiled MCC
1 m
40 μm, 103 
OV-1, 0.2 μm
1700C (100 c)
250C/min up to 2700C
argon, 60 ml/min
Peaks:
1 - alpha-hexachlorocyclohexane;
2 - beta- hexachlorocyclohexane;
3 – hexachlorobenzene;
4 – gamma- hexachlorocyclohexane;
5 – DDE;
6 – DDD;
7 – DDT.

 


Diesel oil

Diesel oil

 

Chromatograph

Detector 
Injection
Column
Length
 Capillaries
LSP
 Temperature

Carrier gas

  TSVET-500M,
upgraded
FID
Splitting 1:1
Coiled MCC
1 m
40 μm, 103
OV-5, 0.05 μm
600C (60 s), 20 0C/min up to 2100C
Ar, 40 ml/min
Peaks:
normal hydrocarbons
C10- C26




Fuel contamination of water

Fuel contamination of water

Water extract (hexane) with adding normal hydrocarbons: C10, C20, C30, C40.
 

Chromatograph

Detector
Injection
Column
Length
 Capillaries
LSP
 Temperature

Carrier gas

  TSVET-500M,
upgraded
FID
Splitting 1:1
Coiled MCC
1 m
40 μm, 103 
OV-1, 0.2 μm
500C (30 s), 20 0C/min up to 3000C
argon, 45 ml/min





Analysis of biological liquids.
Analysis of head space vapors
Analysis of biological liquids

Alcohols
 

Chromatograph

Detector 
Injection
Column
Length
 Capillaries
LSP
 Temperature
Carrier gas

  TSVET-500M,
upgraded
FID
0.5 ml, splitting 1:1
Coiled MCC
1 m
40 μm, 103 
OV-1301, 0.4 μm
400C
Ar, 35 ml/min
Peaks:
1 - ethanol;
2 - acetone;
3 – isopropanol;
4 – propanol;
5 – 2-butanol;
6 – isobutanol;
7 – 1-butanol.



Mud logging

Test mixture of hydrocarbons

Test mixture of hydrocarbons
 

Chromatograph

Detector
Injection
Column
Length
 Capillaries
LSP
 Temperature
Carrier gas

  TSVET-500M,
upgraded
FID
0.5 ml, without splitting
Coiled MCC
1 m
40 μm, 103
OV-1, 0.2 μm
900C
Ar, 40 ml/min
Peaks:
1 – pentane;
2 – hexane;
3 – benzene;
4 – heptane;
5 – toluene;
6 – octane;
7 – n,m-xylenes;
8 – o-xylene;
9 – nonane;
10 – decane.

  © Multichrom Ltd, 2009

Powered by: Amiro.CMS engine