Applications of multicapillary columns in laboratory gas chromatographs
Download file (pdf - 232Kb)
Control of drugs and narcotic substances
 Opium extract |
 |
Chromatograph Detector Injection Column Length Capillaries LSP Temperature Carrier gas |
 |
Kristall 2000M FID Splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-1, 0.2 μm 2000C Ar, 89 ml/min |
Peaks: 1 – methylstearate (standard); 2 – codeine; 3 – morphine; 4 – tebaine; 5 – papaverine; 6 – narcotine
| |
Control of drinking water quality. Analysis of head space vapors

Volatile halogenated hydrocarbons in water
|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature Carrier gas |
 |
Tsvet-500M, upgraded ECD 0.5 ml; splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-1301, 0.4 μm 400C Ar, 65 ml/min |
Peaks: 1 - 1,2-dichloroethane; 2 – chloroform; 3 - carbon tetrachloride; 4 - trichloroethylene; 5 – dichlorobromomethane; 6 - tetrachloroethylene; 7 - dibromochloromethane; 8 – bromoform.
| |
Control of pesticide contaminations

Chloroorganic pesticides.
|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature
Carrier gas |
 |
TSVET-500M, upgraded ECD Splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-1, 0.2 μm 1700C (100 c) 250C/min up to 2700C argon, 60 ml/min |
Peaks: 1 - alpha-hexachlorocyclohexane; 2 - beta- hexachlorocyclohexane; 3 – hexachlorobenzene; 4 – gamma- hexachlorocyclohexane; 5 – DDE; 6 – DDD; 7 – DDT.
| |
Diesel oil

|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature
Carrier gas |
 |
TSVET-500M, upgraded FID Splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-5, 0.05 μm 600C (60 s), 20 0C/min up to 2100C Ar, 40 ml/min |
Peaks: normal hydrocarbons C10- C26
| |
Fuel contamination of water

Water extract (hexane) with adding normal hydrocarbons: C10, C20, C30, C40.
|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature
Carrier gas |
 |
TSVET-500M, upgraded FID Splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-1, 0.2 μm 500C (30 s), 20 0C/min up to 3000C argon, 45 ml/min |
| |
Analysis of biological liquids. Analysis of head space vapors

Alcohols
|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature Carrier gas |
 |
TSVET-500M, upgraded FID 0.5 ml, splitting 1:1 Coiled MCC 1 m 40 μm, 103 OV-1301, 0.4 μm 400C Ar, 35 ml/min |
Peaks: 1 - ethanol; 2 - acetone; 3 – isopropanol; 4 – propanol; 5 – 2-butanol; 6 – isobutanol; 7 – 1-butanol.
| |
Mud logging

Test mixture of hydrocarbons
|
 |
Chromatograph
Detector Injection Column Length Capillaries LSP Temperature Carrier gas |
 |
TSVET-500M, upgraded FID 0.5 ml, without splitting Coiled MCC 1 m 40 μm, 103 OV-1, 0.2 μm 900C Ar, 40 ml/min
|
Peaks: 1 – pentane; 2 – hexane; 3 – benzene; 4 – heptane; 5 – toluene; 6 – octane; 7 – n,m-xylenes; 8 – o-xylene; 9 – nonane; 10 – decane.
| |
|
|